Wednesday, 17 September 2014

qPCR quantification using our new Agilent Bravo robot

We have an exciting new instrument in the Genomics Core which will enable us to automate several of our protocols which until now have been quite labour intensive. This is the Bravo Robot by Agilent. After some previous experience with automation, I think that the results we have seen up until now are really quite promising with the aim of saving hands on time and providing consistency and accuracy within protocols.


qPCR test run - One of the first tests we ran upon installation of the Bravo was qPCR quantification. We quantify all libraries which are submitted to our sequencing service so we can aim to generate cluster densities on the flowcell which will yield large amounts of high quality data.

For this test, a single RNA-seq library was quantified using our standard method with the Illumina quantification kit by Kappa Biosystems. To test the reproducibility of the robot, we performed qPCR on this one sample 24 times as this is the maximum number of tubes which can be loaded per run.

In addition to this, each of the same 24 aliquots of this one library was set up in the same way but manually. This test was useful to determine how good the liquid handling on the Bravo is by looking at the reproducibility of the 24 replicates but additionally for comparing that between manual versus automated set up. The library had also been quantified previously so we expected the concentration to be 50nM.
Agilent Bravo Robot

The results show that there is higher variation in concentrations achieved manually although both methods slightly overestimated the expected concentration of 50nM. We saw an average concentration of 55.6nM manually (an 11% increase from 50nM) in comparison to a concentration of 52.7nM (a 5.5% increase from 50nM) on the Bravo.
Despite there being an about 1.8% difference between the average concentrations seen on the manual set-up in comparison to that of the automated, we can see that the Bravo has yielded more consistant results which is what we would expect and also good news. Since this test, we are now quantifying all SLX library submissions using the Bravo.

Although the robot will not be for general use and we will be unable to run individuals qPCR, we will be using it for all qPCR quantification for libraries submitted to us for our NGS service and for generating RNA-seq libraries. Once these protocols become robust within our lab, we will explore the option of utilising the robot in other protocols including Exome libary prep. It is additionally going to be used for automation of ChIP by the Odom group.


Thursday, 28 August 2014

Help! I really don’t know anything about Next Generation Sequencing…

Over the last three years, while managing the CRUK-CI NGS service, I have heard one phrase more times than I can count:

Help! I really don’t know anything about Next Generation Sequencing…

Yet I have not got bored of hearing this phrase, or of explaining the basics of NGS to fellow researchers and colleagues. That is quite simply because (i) I think NGS is the bee’s knees, and (ii) my answer is never the same twice.

I understand why NGS can be daunting to researchers who wish to use it for the first time; it is technically complex at the same time as being unforgiving if you make a mistake – quite frankly, it is difficult to do well, and expensive if you screw up. The principles underlying the Illumina SBS chemistry are very simple, but the vagaries of instrumentation, the concepts behind designing a good experiment, and the technical skill required to create a high quality library add a layer of complexity which can frighten entire labs away from NGS.

The one thing I always open with is “Don’t be put off; it’s not as complicated as it looks”. The rest of the story changes every month, as new technologies are released and old ones are updated or improved.

Regardless of the name of the current Illumina big-hitter, or the details I advise you take into account when using a particular kit, there are some tips that I give to everyone, which I can also share with you. I will be honest and say they are unashamedly Illumina-focused. I manage an NGS service which runs Illumina sequencers, what else did you honestly expect?

1. Start with the Basics

I have not yet found a better resource for NGS training than the Illumina website (let's overlook the rubbish search function and confusing navigation for the time being). The user-training videos are easy to follow but sufficiently comprehensive to get you started. Furthermore, you can re-watch them once you have some experience, to help with troubleshooting. Don’t even start your experimental design until you have understood how the process works, or you will make expensive mistakes.




2. Practice, Practice, Practice

This should be obvious, but you might be surprised how many folks get it wrong. Don’t ever do your first NGS library preparation with your precious, tender, super-rare, pride-and-joy samples! Don’t even think about it! Don’t expect to buy one 96-sample kit and get 96 high quality NGS libraries with your first prep. Start small. Practice until it works, and then perform your experiment.

When you are ready to try NGS for the first time: buy a small kit, and source a batch of DNA or RNA that you can waste while you learn how to make good libraries. Start by making 8 libraries for practice, following the kit instructions very carefully. Where possible, collect a small quantity of DNA/cDNA after each cleanup step which you could use to troubleshoot failures.

I would strongly recommend that you sequence your best batch of practice libraries, rather than relying solely on the library QC. It is easy to make an RNAseq library which looks quite convincing on the bioanalyzer but performs poorly in differential gene expression analysis, and very, very easy to make a library which looks like a nice small RNA library which contains only the Illumina adapter sequence.

3. It’s Not What You Know, it’s Who You Know

It is unlikely that your first adventure in NGS will be plain sailing, so it’s important to know where you can go for help. If you’re new to NGS and you’re based here at the CRUK Cambridge Institute, the best starting point is to come down to the Genomics Core and talk with us. We’re always happy to meet NGS novices and to help you get your experiments started. If you’re not based in the CRUK CI I’m afraid we can’t offer you much help – we’re a small team and we’re dedicated to supporting this Institute. If you can find a Genomics Facility in your institute then they are going to be your best bet for introductory information, as well as more targeted advice.

If you’re not lucky enough to have a dedicated team onsite then the next best resource for you is Illumina’s technical support team (techsupport@illumina.com). They are well-informed and always happy to help with troubleshooting.

Other great sites to help you troubleshoot your NGS problems are:

Friday, 25 July 2014

PhiX Control - Phact or PhiXion?

Many of our users might have heard us talking about or seen a percentage of their reads aligning to the PhiX genome in the Multi Genome Alignment (see figure below). This is a result of Illumina's recommendation to use the PhiX genome control (see TechNote) for troubleshooting and quality control purposes. There are many features of PhiX that make it a good NGS control: it has a small 5386bp genome, it is a balanced genome (45% GC and 55% AT), and the library is 375bp average making it perfect for clustering and sequencing. The PhiX genome was the first genome to ever be sequenced.

An external file that holds a picture, illustration, etc.
Object name is fgene-05-00031-g001.jpg
The Multi-Genome Alignment report
Why PhiX helps on the sequencer:
We use PhiX control in order to assess the quality of sequencing runs using Sequencing Analysis Viewer or SAV (see image below for an example). Illumina ship PhiX control at 10nM, which we then dilute, denature using our standard protocol and aliquot ready to use prior to clustering. As Illumina suggest, we spike in 1% of PhiX in lanes 1-7 and 5% in lane 8 in all our Hiseq runs, unless requested otherwise. We spike in 5% in our Miseq runs as we see more variable libraries being sequenced here. When checking run performance metrics in SAV, we check the cluster density, clusters passing filter (how many of the clusters are true clusters), error rate, phasing and pre-phasing and also the alignment rate to check the right amount of PhiX spiked in is aligning to the PhiX genome. These results can help us determine whether the problem is associated with the library or the machine, which is the reason we use PhiX to distinguish where the problem lies when a run or a lane doesn't perform well.



PhiX helps our troubleshooting: 
We check the run performance at every stage possible: we check after the clustering step to ensure the fluidics delivery is equal across all lanes, we check first base report to check the run looks good at the start of the sequencing and then we check the run metrics throughout the sequencing. When troubleshooting, we look at the same metrics as mentioned earlier, but in a lot more detail and also at many other metrics such as %base and the images to check that the library is balanced and to check the machine is behaving. When a run hasn't performed as expected and we cannot figure out the cause, we may also get Illumina involved and discuss the runs with them.

It can help get better results with funky libraries: There are many different library prep methods making it difficult to predict the performance of every sequencing run. Some methods such as Bisulfite, iClip, Bless, amplicons etc can produce "funky" libraries that might require the use of up to 50% PhiX.

The Genomics Core recommend using a higher percentage of phiX when:
  • You have a low diversity library (also under clustering can help here as well)
  • You have small amplicon pools
  • Bisulfite-seqeuncing, Blessd, amplicon sequencing

Here is the Illumina product code if you would like to order phiX:
PhiX control v3
FC-110-3001



Thursday, 10 July 2014

Agilent’s Clinical Meeting, Haloplex and SureSelect



Agilent’s Clinical Meeting, Haloplex and SureSelect

I recently attended the Agilent Clinical Meeting in London which gave some very informative presentations on the extent to which Next Generation Sequencing is aiding diagnosis of disease and screening in the clinic. Clinical genetics labs need to be able to provide diagnostic tests which have a very rapid turnaround time that are also cost effective.

Many diagnostic tests are currently based on sequencing a specific gene and using Sanger sequencing. However, many people talked about how they are developing panels for targeted sequencing along with studying the exonic regions by exome sequencing.  Despite the advances of NGS in the clinic, it was clear that whole genome sequencing is where we want to be heading, to get a complete picture. Unfortunately right now, it is not affordable enough.

After attending this conference it was interesting to see how other labs are using Agilent’s Enrichment and panel solutions so thought I would summarise the technologies here.

Haloplex Target Enrichment

What is Haloplex?
HaloPlex is a Target Enrichment System which can be used for the analysis of genomic regions of interest and is aimed for studying a large number of samples.

How does it work?
1.The workflow appears quite simple, starting with DNA fragmentation using restriction enzymes. 

2.Probes which are designed to both ends of a DNA fragment are hybridised to form circular DNA molecules.

3.There is a clean-up step using magnetic streptavidin beads which captures only those fragments containing the biotinylated Haloplex probes. The circular molecules are then closed by ligation.

4.Finally, PCR is used to amplify the targeted fragments ready for sequencing.

What can I enrich for?
SureDesign software can be used to design these custom panels for specific genes or for thousands of exons of interest.

Several clinical labs described how Haloplex technology is enabling them to design diagnostic tests based on screening for specific disease causative genes. Its popularity seemed to be down to its ability to permit a fast turnaround time due to the reduced amount of sample preparation required.

SureSelect Target Enrichment

What is Sureselect?
Agilent’s SureSelect technology enables you to look at the whole exome or at a targeted panel.  It has become a very useful tool in focusing on familial disease loci and for validation of whole genome sequencing. 

How Does it work?
The SureSelect workflow involves a shearing step of gDNA followed by library preparation incorporating adaptors required for sequencing and indexes for multiplexing. Regions of interest are selected for by a 24hour hybridisation step with biotinylated RNA library baits followed by a cleanup step using magnetic streptavidin beads. The baits can be custom designed using Agilent’s SureDesign software. PCR is then used to amplify these regions which are then ready for sequencing.



Exomes in the Genomics Core

Here in the Genomics core, we are currently using Illumina’s Nextera Rapid Exome kit for Exome sequencing and Fluidigm Access Arrays for generating libraries for targeted sequencing.

Agilent have recently released a new SureSelect kit, SureSelectQXT which combines a transposase-based library prep, followed by target enrichment. We have just received one of these kits and will soon be testing this in the lab.


Thursday, 3 July 2014

To seq or not to seq that is the DGE question.

The most common question asked in Differential Gene Expression (DGE) experimental design meetings at the CI is; "should we do RNA-seq or microarray processing?". It all boils down to what questions you want to answer and how the data will integrate into the bigger experiment. I have described some of the most common questions that are asked or discussed and hopefully this information will be useful in getting you thinking about the direction you want to go. 
  1. Why are people doing RNA-seq?
  2. Isn't RNA-seq really expensive?
  3. What about analysis, does it take longer to analyse RNA-seq data?
  4. Do I need as many replicates? 
  5. How long does it take?
  6. How many samples can be processed at once?
1. Why are people doing RNA-seq? RNA-seq data allows you to have a greater dynamic range than microarray. RNA-seq is a digital reading (counting number of reads) and microarray is the analogue reading (fluorescents units) this can be useful if you are looking at the extremes of expression. You may if you wish in the future take your prepared library and do a different type of sequencing and analysis to look for splice junctions and other transcriptional changes. It is important to remember that wanting more than DGE needs a completely different experimental design. For microarray processing you are restrained to the design of the array and what species you want to explore, with RNA-seq there is no restraint as long as you have an adequate reference genome/transcriptome to align your data to. There are lots of technical reasons why you would choose one method over the other but I do not think that you can ignore the fact that RNA-seq is the new technology and people are choosing the method as it fashionable and may seem to be more attractive for publications.

2. Isn’t RNA-seq really expensive? Currently the biggest cost in sequencing is the library preparation. In the core we are currently investigating alternative suppliers to reduce this cost. None the less currently sequencing costs are approximately the same as microarray for DGE analysis within the CIGC.

3. What about analysis, does it take longer to analyse RNA-seq data? By its nature more data is created from RNA-seq sequencing so this in its self requires a significant amount of computing time to process the information. Put these aside similar stringent work flows and pipelines are in place to create a comprehensive gene list of the comparison for both processes.

4. Do I need as many replicates? Yes. The design of the experiment for DGE will remain similar for both RNA-seq and microarray which includes replication requirements. Therefore the number of replicates recommended for the experiment will be the same for either RNA-seq or microarray processing.

5. How long does it take? RNA-seq takes about the same amount of time to process samples in the lab as microarray samples. For both it takes just under a week to get to QC’ed cRNA (microarray) or normalised pooled libraries (RNA-seq). We process both protocols within the institute. However, as we no longer have a working microarray scanner on site, the guys at the Department of Pathology kindly perform the scanning step for us.

6. How many samples can be processed at once? Microarray project designs are constrained to multiples of 12 to get the most out of the consumables, due to the way they are manufactured. RNA-seq utilises 96 individual indexes so if processing less than 94 samples (we use 2 for positive controls) all sample can be pooled together. It gets a little more complicated for larger projects but this is also true for microarray processing.

Friday, 20 June 2014

An Intro to BioNano's Irys System

The BioNano Irys System is an alternative way to look at whole genomes, specifically structural variation. Although it isn't classed as a NGS instrument it has some similarities and could be beneficial when used in conjunction with sequencing. Similar to NGS, the Irys system can be used for genome mapping and de novo assembly. BioNano boasts that the Irys can generate several Gb per hour and can run long sections of DNA at a time.

From the BioNano website
Like NGS, the sample requires some preparation prior to running whereby the DNA is fluorescently labelled. Unlike NGS, this method does not require PCR which eliminates the possibility of PCR bias. Single stranded DNA is used to prevent entwining on itself.

The prepared DNA is fed into Irys Nanochannel which untangles in solution and is moved along the flowcell by electrophoresis. The current is briefly turned off, the DNA is stretched and the machine takes an image of the DNA by exciting the fluorescence. The image is taken with hundreds of thousands of bases of the sequence motifs. This is the data used to create the genome map.

The instrument costs $295,000 which is expensive but not if your used to purchasing NGS instruments.